Selected Publication:
Rottger, A; Honig, A; Schmidt, V; Buchroder, H; Rox, A; Butterweck, G; Schuler, C; Maringer, FJ; Jachs, P; Edelmaier, R; Michielsen, N; Howarth, CB; Miles, JCH; Vargas, A; Ortega, X; Burian, I; Turtiainen, T; Hagberg, N.
(2006):
Radon activity concentration - a Euromet and BIPM supplementary comparison
APPL RADIAT ISOTOPES. APPLIED RADIATION AND ISOTOPES; 64: 1102-1107.
FullText
FullText_BOKU
- Abstract:
- For the first time, a comparison of radon activity concentration in air has been performed within the scope of Euromet. In the project 657, 'Comparison of calibration facilities for the radon activity concentration,' 12 participants from 9 countries compared different radon reference atmospheres at 1, 3 and 10 kBq m(-3) via a transfer standard. The comparison was listed as BIPM supplementary comparison EUROMET.RI(II)-S1. The results of most participants are correlated due to common traceability to one single radon gas standard producer. This makes a careful correlation analysis necessary to achieve an appropriate comparison reference value. The results of the comparison as well as the complex analysis of the correlated set of data is presented and discussed. (c) 2006 Elsevier Ltd. All rights reserved.
- Authors BOKU Wien:
-
Maringer Franz Josef
- Find related publications in this database (Keywords)
-
radon activity concentrations
-
comparison
Altmetric: